Appendix C - Live Animal/Severe Environment Weighing (T)
Note: At any time during the T procedure you can reach the final reading more quickly by
pressing the RANGE/FUNCTION button. This resets the time averaging and allows the balance
to reach the final weight faster.
Procedure
Display Reading
1 Press the RANGE/FUNCTION button. Display cycles repeatedly through PCS, HI-OK-LO, CAL 1,
CAL2,
T,
and %.
2 Press the MODE button when
T
appears on the display.
T
appears in the display and the balance is now in the time
averaging weighing mode.
3 Place a container/cage on the pan.
T
appears in the display as well as a slowly increasing
weight of the container/cage.
4 When the balance has stabilized,
press the ZERO button to zero out the
container/cage weight.
T
appears in the display as well as zeros.
5 Remove the container/cage and place
the sample/animal in the
container/cage.
T
and a slowly decreasing negative weight are displayed.
6 Replace the container/cage with the
enclosed sample/animal on the
weighing pan.
After the sample/caged animal is placed on the pan, the
weight reading will slowly stop decreasing and then slowly
start increasing as the weight reading is averaged over
time. When the weight stops drifting upward, you have
reached the final time averaged weight reading.
7 If you want to reach the final time
averaged weight reading more quickly,
press the RANGE/FUNCTION button
This resets the time averaging mode.
The final time averaged weight reading is reached more
quickly.
8 If you want to exit the time averaging
weighing mode, press the MODE
button.
The balance returns to normal operation.