AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 18 of 22
Appendix A - Front Panel Parts Counting
Procedure
Display Reading
1
Press the RANGE/FUNCTION
button.
Display cycles repeatedly through PCS, HI-OK-LO,
CAL1, CAL2,
T
, and %.
2
Press the MODE button while PCS
appears on the display,
PCS and 0 flash. This is a prompt to zero the balance
with the container you will be using to weigh the initial
sample.
3
Place the empty container on the
pan, wait 10 seconds, then press
the ZERO button.
PCS and 10 flash. This is a prompt to place 10 pieces in
the container and press zero.
4
Place 10 pieces in the container,
wait 10 seconds, then press the
ZERO button.
PCS continues to flash and OK appears on the display
while a stable weight is obtained. Then 10.000, or
something very close to it appears on the display. The
number of decimal places depends on the weight. PCS
continues to flash. This prompts you that you may now
increase the size of the sample if you wish. The decimal
places are an aid. When adding more samples, if the
decimal part approaches .5 you cannot be sure of the
exact count.
5
If you need a larger sample, add
pieces (the balance will count them)
until the decimal part is near 0.7 or
0.3. Then press the ZERO button.
The balance will round to the nearest whole number but
still show the decimal places. This step may be repeated
as many times as you wish.
6
When you are satisfied the sample
is large enough, press the ZERO
button once again.
The balance rounds to the nearest whole number, the
decimal places disappear, and PCS lights steadily.
7
When you wish to count something
else, press the MODE button.
You are now back at step 2 with a flashing PCS and 0.
8
If you want to exit the counting
mode, press the MODE button
again.
You are now in the weighing mode you used just before
selecting pieces.