AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 7 of 22
Connecting the ZSP-150-F to the Vapor Pro
®
/ Vapor Pro Rx
®
The balance is configured to interface directly with the Vapor Pro
®
series of Moisture Analyzers.
Weight readings at the balance can be transmitted directly to the Vapor Pro
®
or Vapor Pro
®
Rx
through the null modem cable connection.
To enable this communication between the balance and the Vapor Pro
®
/ Vapor Pro
®
Rx:
•
Power to both units should be switched OFF.
•
Connect the null modem cable (AZI P/N 200-0069) to the RS-232 connector at the rear of the
balance. Connect the other end of the cable to the balance connector at the rear of the Vapor
Pro
®
/
Vapor Pro
®
Rx.
•
Switch the power to both units ON.
•
On the Vapor Pro
®
/
Vapor Pro
®
Rx, press the
[Menu]
key.
•
Use the
[Down]
key to highlight the SETUP MENU.
•
Press the
[Selct]
key to access this menu.
•
Highlight the EXT BALANCE SETUP option and press
[Selct]
.
•
Highlight and select the SELECT BALANCE DRIVER option.
•
Use the
[Edit]
key to select AZISP150 as the external balance.
•
Press
[Quit]
and then
[Accpt]
key to program this option.
•
Select the CHECK BALANCE COMM option to verify the cable is connected and that the
Vapor Pro
®
/
Vapor Pro
®
Rx and the balance are communicating.
•
Use the
[Esc]
key to return to the MAIN MENU.
If the Communication Check is not successful, the balance’s communication settings need to be reset
to their factory presets. Refer to
RS-232 Front Panel Configuration
on page
12
for instructions on
resetting the balance to the required factory presets.
Using the ZSP-150-F with the Vapor Pro
®
/ Vapor Pro
®
Rx
•
From the MAIN MENU on the connected Vapor Pro
®
/
Vapor Pro
®
Rx, select the MEMORY
START MENU and then ADD/EDIT MEMORY START.
•
Access the memory start to be edited using the
[Up]
or
[Down]
keys and the
[Selct]
key.
•
At the SAMPLE WEIGHT ENTRY MODE option, select DIGITAL BALANCE.
When a test is run using that memory start parameter, the Vapor Pro
®
/
Vapor Pro
®
Rx will
automatically check for data communication. The weight readings from the external balance will be
displayed on the Vapor Pro
®
/
Vapor Pro
®
Rx display. Follow the display prompts to proceed through
the weighing and testing sequence.