![Arizona Instrument ZSP-150-F User Manual Download Page 13](http://html1.mh-extra.com/html/arizona-instrument/zsp-150-f/zsp-150-f_user-manual_2972213013.webp)
AZI Customer Service 800-528-7411 Manual, ZSP-150-F Page 13 of 22
To enter RS-232 configuration mode, proceed as follows:
Procedure
Display Reading
1 Hold the SEND button down while pressing the
ON/OFF button
The current baud rate is displayed
2 Press the SEND button until br 9600 is displayed.
br 9600
3 Press the MODE button to move to the next
parameter: PARITY
The current parity setting, PA EVEn,
PA odd or PA none is displayed.
4 Press the SEND button until PA none is displayed.
PA none
5 Press the MODE button to move to the next
parameter: AUTO SEND
The current auto send setting, AS OFF
or AS on is displayed.
6 Press the SEND button until AS ON is displayed.
AS ON
7 Press the MODE button to move to the next
parameter: CONTINUOUS SEND.
The current continuous send setting,
CS OFF or CS on is displayed.
8 Press the SEND button until CS OFF is displayed.
CS OFF
9 Press the MODE button to move to the next
parameter: SEND WHEN STABLE (SS).
The current send when stable setting,
SS OFF or SS on is displayed.
10 Press the SEND button until SS OFF is displayed.
SS OFF
11 Press the MODE button to move to the next
parameter: DISPLAY BLANKING.
The current display blanking setting, bL
on or bL OFF is displayed.
12 Press the SEND button until bL OFF is displayed.
bL OFF
13 Press the RANGE / FUNCTION button.
SAVEd
is displayed followed by
returning to normal operation.
Changing any of these settings from the AZI factory
presets is NOT recommended, and will affect
communication with the Computrac
®
Vapor Pro
®
moisture analyzers
.
If any of the preset values are changed, use the above instructions
to restore the balance’s ability to communicate with the Computrac
®
Vapor Pro
®
by resetting the values to the preset AZI defaults listed.
Verify the balance’s ability to communicate with the Vapor Pro
®
/ Vapor Pro
®
Rx
as detailed in
Connecting the ZSP-150-F to the Vapor Pro
®
on page
7
.
Settings Summary:
br
9600
PA NONE
AS ON
CS OFF
SS OFF
bL OFF