Ranger Irrigation Fluid Warming System
English
CONDITION
CAUSE
SOLUTION
Alphanumeric display
reads “Er 5” or “Open.”
Electrical interference.
Remove the unit. Refer to
biomedical technician or
call Arizant Healthcare
Inc. Technical Support.
CONDITION
CAUSE
SOLUTION
Alphanumeric display
reads “Er 4” or “Open.”
Open wire on temperature sensor.
Do not use unit. Call
Arizant Healthcare Inc.
Technical Support.
CONDITION
CAUSE
SOLUTION
Alarm indicator light illuminates
and alarm sounds, alphanumeric
display alternately flashes
a temperature of 33°C or
below and the word “LO.”
Under temperature condition
caused by very high flow
using very cold fluid, or
defective heater/relay.
Alarm should stop when
temperature rises above 33°C.
If alarm continues, unplug
unit and discontinue use.
Call Arizant Healthcare
Inc. Technical Support.
CONDITION
CAUSE
SOLUTION
Cannot remove warming
cassette from unit.
Warming cassette is too full,
fluids are still being infused,
or clamp is open proximal
to the warming cassette.
•
Make sure fluid is drained from
warming cassette before sliding
out the cassette, that fluids
are no longer infusing, and
that clamp is closed proximal
to the warming cassette.
•
Warming unit is below
patient level, creating
excessive back pressure.
•
Raise unit above patient level.
•
CONDITION
CAUSE
SOLUTION
Alarm sounds but unit
has been turned off.
Independent backup safety
system has been activated.
Unplug unit. Call
Arizant Healthcare Inc.
Technical Support.
Summary of Contents for Bair Hugger Ranger 247
Page 14: ...Deutsch 13 ...
Page 26: ...Français 27 ...
Page 38: ...Italiano 41 ...
Page 50: ...Español 55 ...
Page 62: ...Português 69 ...
Page 74: ...Nederlands 83 ...
Page 86: ...Dansk 97 ...
Page 98: ...Svenska 111 ...
Page 110: ...Norsk 125 ...
Page 122: ...Suomi 139 ...
Page 134: ...Ελληνικά 153 ...
Page 146: ...日本語 167 ...
Page 157: ...レンジャーイリゲーション液加温システムは 日本語 180 記号の定義 オン オフ ヒューズ 注意 添付文書類を参照してください タイプ B 適用部品 電圧 交番電流 AC ...
Page 158: ...Polski 181 ...
Page 170: ...Magyar 195 ...
Page 182: ...Česky 209 ...