Ranger Irrigation Fluid Warming System
English
Troubleshooting
CONDITION
CAUSE
SOLUTION
Nothing illuminates on the
warming unit panel.
Unit is not turned on, plugged
in, or power cord is not plugged
into an appropriate outlet.
•
Turn unit on. Make sure the
power cord is plugged into
the power entry module of the
warming unit. Make sure the
warming unit is plugged into
a properly grounded outlet.
•
Unit failure.
•
Call Arizant Healthcare
Inc. Technical Support.
•
CONDITION
CAUSE
SOLUTION
Alarm indicator illuminates
and alarm sounds, alpha-
numeric display alternately
flashes a temperature of 48˚C
or higher and the word “HI.”
Temporary over-temperature
condition because:
An extreme change in flow
rates occurred (e.g., from
500 ml/min to stop flow).
•
Open flow to reduce
temperature. Alarms will stop
when the drops below 48°C.
The unit is ready to use.
•
Unit was turned on and reached
set point temperature before
warming cassette was inserted.
•
Alarms will stop when the
temperature drops below 48°C.
The unit is ready to use.
•
Fluids were prewarmed to
above 42°C before being run
through the warming unit.
•
Turn off unit. Discontinue
infusion of fluids. Do not
warm fluids before infusing
them through the Ranger
Irrigation warming unit.
•
CONDITION
CAUSE
SOLUTION
Alarm sounds, alphanumeric
display and alarm indicator
light go dark.
Primary controller failure.
Unit will no longer operate.
Power to heating plates will shut
off if temperature rises to 50˚C.
Turn unit off and unplug it.
Discontinue use of unit. Discard
disposable set. Alarm will continue
to sound if you do not unplug
unit. Call Arizant Healthcare
Inc. Technical Support.
CONDITION
CAUSE
SOLUTION
Unit alarms soon after
plugging it in (unit does not
have to be turned on for
this condition to occur).
Heater temperature rises to
50˚C and unit shuts down
soon after plugging it in (unit
does not have to be turned on
for this condition to occur).
•
•
Test screw on bottom of
unit is loose or missing.
Make sure test screw is completely
tightened. If it is missing,
turn unit off and unplug
it. Call Arizant Healthcare
Inc. Technical Support.
Summary of Contents for Bair Hugger Ranger 247
Page 14: ...Deutsch 13 ...
Page 26: ...Français 27 ...
Page 38: ...Italiano 41 ...
Page 50: ...Español 55 ...
Page 62: ...Português 69 ...
Page 74: ...Nederlands 83 ...
Page 86: ...Dansk 97 ...
Page 98: ...Svenska 111 ...
Page 110: ...Norsk 125 ...
Page 122: ...Suomi 139 ...
Page 134: ...Ελληνικά 153 ...
Page 146: ...日本語 167 ...
Page 157: ...レンジャーイリゲーション液加温システムは 日本語 180 記号の定義 オン オフ ヒューズ 注意 添付文書類を参照してください タイプ B 適用部品 電圧 交番電流 AC ...
Page 158: ...Polski 181 ...
Page 170: ...Magyar 195 ...
Page 182: ...Česky 209 ...