Alert E32: TDis Sens Err –
Refrigeration unit alert.
•
Alert E33: TSuc Sens Err –
Refrigeration unit alert.
•
Alert E30: TAux Sens Err –
Refrigeration unit alert.
•
Alert A2: LP Alarm –
Refrigeration unit alert.
•
Alert A11: No Rfg. Sel –
Refrigeration unit alert.
•
Alert A96: Max Dis. Temp –
Refrigeration unit alert.
•
Alert A16: DI2 Alarm –
Refrigeration unit alert.
•
Alert A17: HP Alarm –
Refrigeration unit alert.
•
Alert A97: Safety Alarm –
Refrigeration unit alert.
•
Alert A45: Standby Mode –
Refrigeration unit alert.
•
Alert A80: Cond. Blocked –
Refrigeration unit alert.
•
High CO2 level (input)
–
High CO2 level (digital input) inside the chamber (only
available in models with CO2 detector). For safety reasons this alarm is verified even
if “CHAMBER OFF”;
•
High CO2 level –
CO2 level above the maximum set value, during the time set for
alarm delay;
•
Low CO2 level –
CO2 level below the maximum set value, during the time set for
alarm delay;
•
CO2 level band –
CO2 level above the value obtained by adding the setpoint with
the band, or below the value obtained by subtracting the setpoint with the band,
during the time set for alarm delay;
•
Evaporator 2 temperature –
Evaporator 2 temperature above 70ºC. The alarm
stops when the temperature of the evaporator 2 register a value below 69ºC;
•
Lamps cut temperature –
Lamps temperature above 50ºC. The alarm stops when
the temperature of the lamps register a value below 45ºC;
•
Lamps limit temperature –
Lamps temperature above 45ºC. The alarm stops when
the temperature of the lamps register a value below 40ºC;
Summary of Contents for FITOCLIMA 2500 HP
Page 1: ...USER S MANUAL READ AND KEEP THIS INSTRUCTIONS ORIGINAL MANUAL EN ...
Page 3: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 3 ...
Page 13: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 13 Top view ...
Page 14: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 14 Side view ...
Page 15: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 15 3D view ...
Page 39: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 39 4 4 Set values configuration ...