FITOCLIMA 2500 HP
Mod. MI 271 EN / R00 23-10-2015
-44-
To change the parameters, simply press on the desired value, enter the new value using the
numeric keypad and confirm with the Enter key.
Note: On some equipment models additional temperature and humidity sensors are installed,
to allow comparison between them and detect any possible sensor failure. The fault is
detected by the difference between the sensors been greater than the set band, for more than
10 minutes, and an alarm is generated to indicate the alarm situation.
Note: The maximum and minimum security temperatures ensures in case of failure the
chamber’s stop, and the activation of an visible/audible alarm, to avoid a possible
overheating or overcooling of the chamber and material under test.
The maximum security temperature should have an higher value than the maximum
temperature programmed in the controller. The same way, the minimum security
temperature should have a lower value than the minimum temperature programmed in the
controller.
All alarm settings must always be properly adjusted to the test.