4.9
Program evolution
The program’s evolution is defined by Ramp ON
/
Ramp OFF:
•
If Ramp OFF each segment defines a Set-Point during the segment time.
•
If Ramp ON, evolution will be in ramp during the segment time.
Relative Humidity will evolve in ramp except when it comes from 0 (OFF) or when it goes
to OFF, that is., if in a certain segment the Relative Humidity Set-Point is 0 (OFF), even if
Ramp ON the Relative Humidity will go immediately OFF, if in a previous Relative
Humidity segment was OFF and the segment taking place is, for example 50%rH, the
Relative Humidity immediately starts controlling to 50%rH. Radiation intensity and CO2
level has similar evolution to Humidity..
Example of a program:
Segment 0:
T=15ºC, H=50%rH, 2 min
Segment 1:
T=30ºC, H=0%rH, 5 min
Segment 2:
T=30ºC, H=50%rH, 3 min
Segment 3:
T=15ºC, H=80%rH, 5 min
Segment 4:
T=15ºC, H=80%rH, 10 min
Segment 5:
T=30ºC, H=50%rH, 3 min
Segment 6:
T=30ºC, H=50%rH, 0 min
(
PG-END
)
Ramp ON
Ramp OFF