FITOCLIMA 2500 HP
Mod. MI 271 EN / R00 23-10-2015
-32-
Section 3
Safety thermostat
The protection function for maximum and minimum temperatures are implemented in the
auxiliary control board, which provides, in case of abnormality, stopping the chamber and
activating an audible alarm, which prevents possible overheating or overcooling of the
interior the chamber and the material under test.
The maximum temperature thermostat must be adjusted to a temperature considerably above
the maximum programmed into the controller, while the minimum temperature thermostat
must be adjusted to a temperature well below the minimum programmed into the controller.
Example:
• High temperature alarm: +30 º C.
The maximum security temperature should be set to +40 ° C.
• Low temperature alarm: +20 º C.
The minimum security temperature should be set to +10 º C.
To change the parameters see "Alarms configuration" chapter under controller section.
All alarm settings must always be properly adjusted to the test.