FITOCLIMA 2500 HP
Mod. MI 271 EN / R00 23-10-2015
-49-
4.10
Program edit
To create or edit a program simply select the desired program and select “Edit”. At the edit
screen press on the desired vales to be changed, modify them, and press Enter to validate.
To save the values press [UP]/[DOWN] or [BACK].
Note: although the programmed time and set values are independent for each segment, the
options “Ramp”, “Repeat” and “Next Program” are common to all segments, in other words
they are program parameters, not segment specific parameters.
Note: to navigate between segments press [UP]/[DOWN] keys to move to the next/previous
segment.
4.11
Program execution
If a program is running (PG-RUN, PG-HLD or PG-END), simply press on the icon at the
title of the main screen, to access the information about the its current state.
The elapsed time of the current segment is changeable. If placed at the maximum value of
the segment, allows to jump to the next segment. Its main function is to allow an adjustment
of program execution with the current time or skip any segments. The maximum allowed
value is defined in the current segment. To change the value simply press on it, modify it,
and press Enter to validate.
The segment execution is sequential, that is the first segment to run is the segment 0,
followed by the segment 1, 2, etc.
Summary of Contents for FITOCLIMA 2500 HP
Page 1: ...USER S MANUAL READ AND KEEP THIS INSTRUCTIONS ORIGINAL MANUAL EN ...
Page 3: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 3 ...
Page 13: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 13 Top view ...
Page 14: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 14 Side view ...
Page 15: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 15 3D view ...
Page 39: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 39 4 4 Set values configuration ...