13
2.5
Selecting the Appropriate Wire Gauge
Below is a table which provides recommended appropriate wire gauges for the
300XAC.
310XAC
Input
Output
Terminal
Line
Neutral
Ground
Line
Neutral
Ground
Nsense
Lsense
Gauge
12 AWG
12 AWG
12 AWG
14 AWG
14 AWG
14 AWG
18 AWG
18 AWG
320XAC
Input
Output
Terminal
Line
Neutral
Ground
Line
Neutral
Ground
Nsense
Lsense
Gauge
12 AWG
12 AWG
12 AWG
14 AWG
14 AWG
14 AWG
18 AWG
18 AWG
340XAC
Input
Output
Terminal
Line
Neutral
Ground
Line
Neutral
Ground
Nsense
Lsense
Gauge
10 AWG
10 AWG
10 AWG
10 AWG
10 AWG
10 AWG
18 AWG
18 AWG
360XAC
Input
Output
Terminal
Line(A,B,C)
Neutral (A,B,C)
Ground
Line
Neutral
Ground
Nsense
Lsense
1Φ2W
10 AWG
10 AWG
10 AWG
8 AWG
8 AWG
8 AWG
18 AWG
18 AWG
3Φ3W
12 AWG
N/A
12 AWG
3Φ4W
14 AWG
14 AWG
14 AWG
2.6
Power Cable
Before connecting power to this instrument, the protective
ground (earth) terminals of this instrument must be
connected to the protective conductor of the line (mains)
power cord. The main plug shall only be inserted in a socket outlet (receptacle)
provided with a protective ground (earth) contact. This protective ground (earth)
must not be defeated by the use of an extension cord without a protective
conductor (grounding).
The mains plug is used as the disconnecting device and
shall remain readily operable. The socket-outlet shall be
installed. Near the equipment and shall be easily
accessible.
The main plug shall only be inserted in a socket outlet with
a protective ground (earth) contact. This protective ground
must not be defeated
by the use of an extension cord
without a protective conductor.
Summary of Contents for 00310XAC
Page 16: ...10 310XAC 320XAC...
Page 24: ...18 Accuracy L 0 1 Hz 0 0 500 Hz H 0 2 Hz 501 1000 Hz...
Page 26: ...20 A...
Page 36: ...30...
Page 101: ...95 move back to the test screen...
Page 108: ...102...
Page 127: ...121 CONNECT CONNECT 0 1 0 1...
Page 130: ...124 Register...
Page 138: ...132...