115
RESET
Turns the output voltage off or resets the instrument in the event of a failure.
TD?
Read the active data being displayed on the LCD display while the test is in
process. Will also read the last data taken when the test sequence has
completed. Each parameter is separated by commas and includes memory
number, step number, test status, frequency value, voltage value, current value,
power value, peak current value, power factor value and timer metering. The
syntax for the command response is {memory, step, status, frequency, voltage,
current, power, peak current, power factor, timer}. Each meter will contain only
the value and not the units. Current and peak current are displayed in amps while
power is displayed in Watts.
RD <step number>?
Read the results for an individual step. The step number is the actual step
number that has been saved within the file, not the order of which the steps were
executed. For example if the test was executed starting from step 3 and ending
with step 5 then the first step test results will be found in location 3 not in location
1. Each parameter is separated by commas and includes step number, test type,
test status, and metering. The syntax for this command response is {memory,
step, status, frequency, voltage, current, power, peak current, power factor, timer}.
ACW test displays 4 meters. Each meter will contain only the value and not the
units. Each meter will contain only the value and not the units. Current and peak
current are displayed in amps while power is displayed in Watts.
TDFREQ?
Read the active frequency value being displayed while a test is in process.
TDVOLT?
Read the active voltage value being displayed while a test is in process.
TDCURR?
Read the active current value being displayed while a test is in process.
TDAP?
Read the active peak current value being displayed while a test is in process.
TDP?
Read the active power value being displayed while a test is in process.
TDPF?
Read the active power factor value being displayed while a test is in process.
Summary of Contents for 00310XAC
Page 16: ...10 310XAC 320XAC...
Page 24: ...18 Accuracy L 0 1 Hz 0 0 500 Hz H 0 2 Hz 501 1000 Hz...
Page 26: ...20 A...
Page 36: ...30...
Page 101: ...95 move back to the test screen...
Page 108: ...102...
Page 127: ...121 CONNECT CONNECT 0 1 0 1...
Page 130: ...124 Register...
Page 138: ...132...