135
Enter the voltage reading from the voltmeter with the numeric keypad. When the
value has been selected press the Enter soft key and you will be moved to the
next calibration parameter Current xx.xA. If you press the Esc soft key you will
be returned to the calibration mode screen.
10.14
Calibration of High and Low Current Range
Use the up or down arrow soft keys to navigate to the Current x.xxxA, or Current
xx.xxA parameter and press the Select soft key.
Follow the prompt message provided on the display, and press the Test/Reset
button to move into the calibration screen for current. If you press the Exit soft
key at this screen you return to the calibration mode screen.
Enter the current reading from the ammeter with the numeric keypad. When the
value has been selected press the Enter soft key and you will be moved to the
next calibration parameter. The soft keys Volt+ and Volt- are available, if needed,
to adjust the voltage output of the instrument to set a proper current value if a
non-recommended load is used. If you press the Esc soft key you will be
returned to the calibration mode screen.
10.15
Calibration of High and Low Power Range
Use the
,
soft keys to navigate to Power xx.xW, or Power xxxxW parameter
and press the Select soft key.