SECTION 7: CONNECTORS, APPLICATIONS & PROTECTION FEATURES
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
42L Series Modular DC Load Operation Manual
Page 52 of 88
KEY# Description
2.
OCP test function parameter settings. There are four parameters for the OCP test function.
These parameters are: ISTAR, ISTEP, ISTOP and Vth. Press the “OCP” key again to set the
OCP test parameter ISTAR(start current point) Press the “OCP” key again to proceed to the
next parameter in order of ISTEP, ISTOP, Vth and Disable. Press any another key to exit the
and save the OCP settings.
The OCP test parameter descriptions are as follows:
a.
ISTAR: Setting the start current point. The LCD display shows “OCP”, “ISTAR” and
0.000A (or initial setup value) from top to bottom on the 5-digit LCD displays. The
setting range is 0.000A to the full scale of the CC mode specification. This
parameter can be adjusted by turning the rotary knob.