INSTALLATION AND HARDWARE MANUAL
ADI 1075 Pilot System
BASIC UNIT
December 2002
2-10
The internal power section has its
own fuses (the image shows the
internal power supply board that is
located at the transformer at the
bottom of the ADI 1075 cabinet):
F1 = Main Fuse,
F2 = Fuse circuit 24 Vdc,
F3 = Fuse circuit 12 Vac,
F4 = Fuse circuit 230 Vac.
Fuse type: 5 x 20 mm.
Fuse rating:
F1 =
T 5A (110/120 Vac),
T 2.5A (230 Vac),
F2 =
T 6.3A,
F3 =
T 2A,
F4 =
T 1A.
Note:
The 24 Vdc circuit is used for actuator supply and supply for the thermal cut-off of the
heater of the Wet Part.
The 12 Vdc circuit is used for illumination assembly supply (internal illumination of the
Bioreactor).
The 230 Vac circuit is used for circulation pump supply (Wet Part).