INSTALLATION AND HARDWARE MANUAL
ADI 1075 Pilot System
BASIC UNIT
December 2002
2-8
MAINS SUPPLY: 400V3~ or 200V3~ High Voltage Current
The ADI 1075 Pilot System is supplied with a voltage selection package; this package is
available in two voltage versions:
Z310750020
: Pilot System voltage selection package 115 Vac,
Z310750030
: Pilot System voltage selection package 230 Vac.
Verify if the proper voltage selection package was installed (see voltage sticker at the main
supply cable gland); if this is not correct, contact your service office.
CAUTION! ELECTRICAL SHOCK HAZARD!
Connect the main supply cable to the power source according to:
- brown (bn)
= L1,
- black1 (bk1)
= L2,
- black2 (bk2)
= L3,
- blue (bl)
= neutral,
- green/yellow (gn/ye)
= earth.
Caution:
Minimum specification of the mains cable = 5 x 2.5 mm
2
. An external mains
supply switch (4 poles, 25 A) must be installed in order to switch off the ADI
1075 Pilot System completely (in case of service).
This external switch must be marked as “disconnecting device for the ADI
1075 Pilot System”.
Using an earth of poor quality may cause malfunctioning of the control
equipment!
Internally each of the three live circuits is used as a separate 115/230 Vac circuit.