background image

- 13 -

4) 

Auto. Power off (P04)

 – The auto. power off time can be set to 10 to 20 minutes. For example, if 

15 minutes is set, the meter will automatically shut down after 15 minutes if no operation; when 
“Off” is set, the auto. power off function will be turned off. Long press 

 to manually shut down 

the meter.

5) 

pH Calibration Reminder (P07) and Conductivity Calibration Reminder (P11)

 – set XX hours (H) 

Or XX days (D) in ZenTest mobile app – settings – Parameter – pH – Calibration Reminder. On 
the meter, you can only check the values that’s been set up on ZenTest App. For example, if 3 
days is set up, the Er6 icon (see Figure-4) will appear in the lower right corner of the LCD screen 
in 3 days to remind you to perform calibration, also in the ZenTest App there will be a pop-up 
reminder. After calibration is finished or the reminder setting is cancelled in the ZenTest App, the 
Er6 icon will disappear.

6) 

pH Back to Factory Default (P08) and Conductivity Back to Factory Default (P12)

 – Select “Yes” 

to set the meter to its default status (erase all calibration record and return all parameter settings 
to the default value). This function can be used when the meter does not work well in calibration 
or measurement. Calibrate the meter again after setting the meter to factory default.  

14. Technical Specifications

pH

Range

-2.00 to 16.00 pH

Resolution

0.01 pH

Accuracy

±0.01 pH ±1 digit

Calibration Points

1 to 3 points

Auto. Temperature 

Compensation

0 – 50˚C (32 – 122˚F)

Conductivity

Range

0 to 199.9 μS, 200 to 1999 μS, 2 to 20.00 mS

Resolution

0.1/1 μS, 0.01 mS

Accuracy

±1% F.S

Calibration Points

1 to 3 points

TDS

Range

0.1 ppm to 10.00 ppt

TDS Factor

0.40 to 1.00

Salinity

Range

0 to 10.00 ppt

Resistivity

Range

50Ω to 20MΩ

ORP (mV)

Range

-1000 mV to 1000 mV

Accuracy

±0.2% F.S

Temperature

Range

0 to 50˚C (32-122˚F)

Accuracy

±0.5˚C

Summary of Contents for ZenTest PC60-Z

Page 1: ...1 PC60 Z Smart Water Tester pH Cond TDS Salinity Resistivity ORP Temp User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Page 2: ...ure below Users can remove the shield when cleaning the sensor and put it back on after cleaning 1 Introduction 3 2 What s in the Kit 4 3 Keypad Functions 4 4 Preparation before Use 5 5 Battery Replac...

Page 3: ...lable on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mo...

Page 4: ...ong press to turn off the tester short press to turn on off backlight 1 In measurement mode short press to switch parameter pH Cond TDS Sal Res 2 In measurement mode long press to turn on off Bluetoot...

Page 5: ...ionized water RO water or tap water is the alternative then shake off excess water 5 3 Perform calibration For pH calibration tutorial refer to Section 6 for conductivity calibration refer to Section...

Page 6: ...tely after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH...

Page 7: ...eck whether pH probe is damaged or broken If so please replace with a new one 2 T h e g l a s s b u l b o r j u n c t i o n i s s e ve re l y contaminated Please use a soft brush with soap water to cl...

Page 8: ...at info aperainst com to find the specialized meter designed for pure water pH test 8 Conductivity Calibration 8 1 How to calibrate 8 1 1 Power on the tester Press to switch to conductivity Cond meas...

Page 9: ...The meter cannot recognize the conductivity standard solutions 1 Make sure the probe is fully submerged in the solution 2 Check if the standard solution is expired or contaminated 3 Check if the condu...

Page 10: ...1XXX ppm c The TDS and Salinity values are converted from the conductivity values via a certain conversion factor TDS and conductivity is linear related and its conversion factor is 0 40 to 1 00 Adju...

Page 11: ...vity sensor 3 If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes over time by its nature Just rinse them off and keep...

Page 12: ...the meter is turned off long press to enter parameter setting short press to switch P01 P02 P14 Short Press parameter flashes short press to adjust parameter short press to confirm Short press to exit...

Page 13: ...or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to set the meter to its def...

Page 14: ...the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing The replacement probes that are compatible with PC60 Z PC60 DE...

Page 15: ...o 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible damage of the probe and it s within th...

Page 16: ...d void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits a...

Reviews: