- 9 -

8.2  Conductivity Calibration Notes
     The tester can calibrate 84μS, 1413 μS and 12.88 mS conductivity calibration solution.  User can 

conduct 1 to 3 points calibration. Refer to the table below.

    
     For conductivity calibration solutions, we recommend replacing the solutions after up to 5 times 

of use to keep the standard solution’s accuracy. Do NOT pour the used calibration solutions back 
into the solution bottles in case of contamination.

       * Once conuductivity is calibrated, TDS, saliniity, and resistivity will be automatically calibrated.

Calibration Indication Icon

Calibration Standards

Measuring Range

84 μS

0 - 200 μS

1413 μS

200 - 2000 μS

12.88 mS

2 - 20 mS

8.3  For the self-diagnosis information, please refer to the table below:

Symbol

Self-Diagnosis information

How to fix

The meter cannot recognize the 

conductivity standard solutions.

1. Make sure the probe is fully submerged in the 

solution.

2. Check  if  the  standard  solution  is  expired  or 

contaminated.

3. Check if the conductivity probe (two black rods) is 

damaged.

4. Check if the conductivity probe is contaminated. 

If so, please use a soft brush with warm water to 

clean up.

 is pressed before 

measurement is 

fully stabilized.

Wait for   to come up and stays on screen

 before pressing 

 

During calibration, readings 

being unstable for over 3 

minutes

1. Shake the probe to remove the air bubbles on the 

surface of the black rods

2. Check if the conductivity probe is contaminated. 

If so, please use a soft brush with warm water to 

clean up.

3. Soak  the  probe  in  12.88  mS  solution  for  10 

minutes, then rinse with distilled water.

The calibration reminder is 

triggered. It’s time to perform a 

new conductivity calibration.

Perform conductivity calibration or cancel calibration 

reminder in ZenTest settings.

Summary of Contents for ZenTest PC60-Z

Page 1: ...1 PC60 Z Smart Water Tester pH Cond TDS Salinity Resistivity ORP Temp User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Page 2: ...ure below Users can remove the shield when cleaning the sensor and put it back on after cleaning 1 Introduction 3 2 What s in the Kit 4 3 Keypad Functions 4 4 Preparation before Use 5 5 Battery Replac...

Page 3: ...lable on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mo...

Page 4: ...ong press to turn off the tester short press to turn on off backlight 1 In measurement mode short press to switch parameter pH Cond TDS Sal Res 2 In measurement mode long press to turn on off Bluetoot...

Page 5: ...ionized water RO water or tap water is the alternative then shake off excess water 5 3 Perform calibration For pH calibration tutorial refer to Section 6 for conductivity calibration refer to Section...

Page 6: ...tely after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH...

Page 7: ...eck whether pH probe is damaged or broken If so please replace with a new one 2 T h e g l a s s b u l b o r j u n c t i o n i s s e ve re l y contaminated Please use a soft brush with soap water to cl...

Page 8: ...at info aperainst com to find the specialized meter designed for pure water pH test 8 Conductivity Calibration 8 1 How to calibrate 8 1 1 Power on the tester Press to switch to conductivity Cond meas...

Page 9: ...The meter cannot recognize the conductivity standard solutions 1 Make sure the probe is fully submerged in the solution 2 Check if the standard solution is expired or contaminated 3 Check if the condu...

Page 10: ...1XXX ppm c The TDS and Salinity values are converted from the conductivity values via a certain conversion factor TDS and conductivity is linear related and its conversion factor is 0 40 to 1 00 Adju...

Page 11: ...vity sensor 3 If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes over time by its nature Just rinse them off and keep...

Page 12: ...the meter is turned off long press to enter parameter setting short press to switch P01 P02 P14 Short Press parameter flashes short press to adjust parameter short press to confirm Short press to exit...

Page 13: ...or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to set the meter to its def...

Page 14: ...the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing The replacement probes that are compatible with PC60 Z PC60 DE...

Page 15: ...o 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible damage of the probe and it s within th...

Page 16: ...d void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits a...

Reviews: