9. Conductivity Measurement

9.1  How to Measure Conductivity, TDS, Salinity, and Resistivity
      Short press 

 to turn on the tester. Rinse the probe in pure water, shake off excess water. Insert 

the probe in your sample solution, make a quick stir and hold still. Record the reading when it is 
stabilized

 

 comes up and stays on screen). Press 

 to switch from conductivity to TDS, 

salinity, and resistivity. Or if you turn on the Auto-Hold function (refer to Section 13.2 (2)), the 
reading will be automatically locked when it's stable. Short press 

 to cancel the Auto-Hold and 

keep measuring.

9.2  Unit Conversion
a.  1000 µS = 1 mS = 1 EC (In conductivity mode, the unit will automatically turn from µS to mS if the 

reading is greater than 1999 µS, meaning you will only see 2.XX mS instead of 2XXX µS)

b.  1000 ppm = 1 ppt (In TDS mode, the unit will automatically turn from ppm to ppt if the reading is 

greater than 999 ppm, meaning you will only see 1.XX ppt instead of 1XXX ppm)

c.  The TDS and Salinity values are converted from the conductivity values via a certain conversion 

factor. TDS and conductivity is linear related, and its conversion factor is 0.40 to 1.00. Adjust the 
factor in parameter setting P13 based on the requirements in different industries. The factory 
default setting is 0.71.

d.  Salinity and conductivity is linear related, and its conversion factor is 0.5. 
e.  Conversion Example: if conductivity measurement is 1000µS, then the default TDS measurement 

will be 710 ppm (under the default 0.71 conversion factor), and the salinity be 0.5 ppt. If TDS 
conversion rate is changed to 0.5, then the TDS measurement will be 500 ppm.

9.3  Temperature compensation factor
      The default setting of the temp. compensation factor is 2.0%/˚C. User can adjust the factor based 

on test solution and experimental data in parameter setting P10. The following table is some 
common examples for setting up the temp. compensation factor.

Solution

Temperature 

compensation factor

Solution

Temperature 

compensation factor

NaCl

2.12%/˚C

10% Hydrochloric acid

1.32%/˚C

5% NaOH

1.72%/˚C

5% Sulfuric acid

0.96%/˚C

Dilute ammonia

1.88%/˚C

Summary of Contents for ZenTest PC60-Z

Page 1: ...1 PC60 Z Smart Water Tester pH Cond TDS Salinity Resistivity ORP Temp User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Page 2: ...ure below Users can remove the shield when cleaning the sensor and put it back on after cleaning 1 Introduction 3 2 What s in the Kit 4 3 Keypad Functions 4 4 Preparation before Use 5 5 Battery Replac...

Page 3: ...lable on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mo...

Page 4: ...ong press to turn off the tester short press to turn on off backlight 1 In measurement mode short press to switch parameter pH Cond TDS Sal Res 2 In measurement mode long press to turn on off Bluetoot...

Page 5: ...ionized water RO water or tap water is the alternative then shake off excess water 5 3 Perform calibration For pH calibration tutorial refer to Section 6 for conductivity calibration refer to Section...

Page 6: ...tely after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH...

Page 7: ...eck whether pH probe is damaged or broken If so please replace with a new one 2 T h e g l a s s b u l b o r j u n c t i o n i s s e ve re l y contaminated Please use a soft brush with soap water to cl...

Page 8: ...at info aperainst com to find the specialized meter designed for pure water pH test 8 Conductivity Calibration 8 1 How to calibrate 8 1 1 Power on the tester Press to switch to conductivity Cond meas...

Page 9: ...The meter cannot recognize the conductivity standard solutions 1 Make sure the probe is fully submerged in the solution 2 Check if the standard solution is expired or contaminated 3 Check if the condu...

Page 10: ...1XXX ppm c The TDS and Salinity values are converted from the conductivity values via a certain conversion factor TDS and conductivity is linear related and its conversion factor is 0 40 to 1 00 Adju...

Page 11: ...vity sensor 3 If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes over time by its nature Just rinse them off and keep...

Page 12: ...the meter is turned off long press to enter parameter setting short press to switch P01 P02 P14 Short Press parameter flashes short press to adjust parameter short press to confirm Short press to exit...

Page 13: ...or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to set the meter to its def...

Page 14: ...the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing The replacement probes that are compatible with PC60 Z PC60 DE...

Page 15: ...o 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible damage of the probe and it s within th...

Page 16: ...d void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits a...

Reviews: