background image

- 11 -

11. Probe Cleaning

1)  The tester is only as accurate as the probe is clean. Always thoroughly rinse off the probe before  

and after each measurement with pure water in a container or with a wash bottle. 

2)  For tough contaminants, detach the sensor shield, soak the probe in Apera probe cleaning solution 

(AI1166) or detergent water for 30 minutes. Then use a soft brush to remove the contaminants. 
Afterwards, soak the probe in 3M KCL soaking solution for at least 1 hour. Rinse it off, then re-
calibrate the tester before using again.

12. Probe Storage

1)  For regular storage, just make sure there are a few water droplets inside the probe cap. Then 

close on the probe cap tightly to maintain the humidity inside so the probe will not dry out.

2)  If the probe is dried out by accident or if you find the probe's response turns much slower than 

usual, soak the probe in the 3M KCL soaking solution for about 2 hours to recover its sensitivity. 
NEVER leave the probe in the 3M KCL soaking solution for longer than 24 hours as it may cause 
damage to the conductivity sensor.

3)  If you find white crystals inside or outside the probe cap, it is perfectly normal. It is the 3M KCL 

soaking solution that crystalizes over time by its nature. Just rinse them off and keep using. This 
chemical is not poisonous nor dangerous, and the probe’s performance will not be affected at all.

4)  NEVER store the probe in pure water like tap, RO, distilled, or deionized water as they could 

damage the pH probe. If this happens, immediately soak the pH probe in the 3M KCL soaking 
solution overnight, then re-calibrate it before using. Pure water is only for rinsing the probe.

10. ORP Measurement

1)  ORP stands for Oxidation-Reduction Potential, measured in mV. It’s also called redox. ORP is a 

measure of the cleanliness of water & its ability to break down contaminants. A separate ORP 
probe (ORP60-DA) needs to be installed to be able to measure ORP. 

2)  Power on the PC60-Z tester, unscrew the original probe, and install the ORP60-DA probe, then 

the tester will automatically switch to ORP measurement mode (Refer to Section 16 for how to 
replace a probe).

3)  Rinse the probe in pure water and shake off excess water. Dip the probe in sample solution, make 

a quick stir and hold still. Record the ORP reading after it's fully stabilized   stays on screen.

Summary of Contents for ZenTest PC60-Z

Page 1: ...1 PC60 Z Smart Water Tester pH Cond TDS Salinity Resistivity ORP Temp User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Page 2: ...ure below Users can remove the shield when cleaning the sensor and put it back on after cleaning 1 Introduction 3 2 What s in the Kit 4 3 Keypad Functions 4 4 Preparation before Use 5 5 Battery Replac...

Page 3: ...lable on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mo...

Page 4: ...ong press to turn off the tester short press to turn on off backlight 1 In measurement mode short press to switch parameter pH Cond TDS Sal Res 2 In measurement mode long press to turn on off Bluetoot...

Page 5: ...ionized water RO water or tap water is the alternative then shake off excess water 5 3 Perform calibration For pH calibration tutorial refer to Section 6 for conductivity calibration refer to Section...

Page 6: ...tely after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH...

Page 7: ...eck whether pH probe is damaged or broken If so please replace with a new one 2 T h e g l a s s b u l b o r j u n c t i o n i s s e ve re l y contaminated Please use a soft brush with soap water to cl...

Page 8: ...at info aperainst com to find the specialized meter designed for pure water pH test 8 Conductivity Calibration 8 1 How to calibrate 8 1 1 Power on the tester Press to switch to conductivity Cond meas...

Page 9: ...The meter cannot recognize the conductivity standard solutions 1 Make sure the probe is fully submerged in the solution 2 Check if the standard solution is expired or contaminated 3 Check if the condu...

Page 10: ...1XXX ppm c The TDS and Salinity values are converted from the conductivity values via a certain conversion factor TDS and conductivity is linear related and its conversion factor is 0 40 to 1 00 Adju...

Page 11: ...vity sensor 3 If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes over time by its nature Just rinse them off and keep...

Page 12: ...the meter is turned off long press to enter parameter setting short press to switch P01 P02 P14 Short Press parameter flashes short press to adjust parameter short press to confirm Short press to exit...

Page 13: ...or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to set the meter to its def...

Page 14: ...the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing The replacement probes that are compatible with PC60 Z PC60 DE...

Page 15: ...o 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible damage of the probe and it s within th...

Page 16: ...d void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits a...

Reviews: