98
211_0010 - ROXY Exceed user manual - rev 04
In stand-alone mode the valve position can be controlled via the F5 func-
tion button in the STAT screen. In the case of a manual valve the infor-
mation above the F5 button (INJ=L) in the STAT screen shows the status
(position) of the valve only, no valve control is possible with this type of
valve.
Fig. 48. Example of the valve flow path of a 6-port valve.
In Dialogue Elite the valve status/position is shown under the monitor tab.
To make it visible in the left table, check the valve position checkbox in the
item selector (by default the valve position is not show).
Make sure that the valve flow path is connected correctly and LOAD
corresponds with position A and INJECT with position B. See example
in the figure below for a 6-port valve.