CHAPTER 9
Rear panel I/O
101
Fig. 51. Top: CONFIG screen. Bottom: ROXY Exceed signal processing
from electrochemical flow cell to output. R is a selectable I/E resistor of
100M
Ω, 10MΩ, 100KΩ, 1KΩ, 100 Ω and 10 Ω in the I/E converter circuit.
The output parameter Vout source can also be selected/set using Dialogue
Exceed. Go to the device settings menu and Acquisition tab and set the
analogue output to DAC or I/E converter. The instrument is by default set
to DAC. For regular measurements it is advised to use the DAC output.
Fig. 52. Dialogue Elite Acquisition tab analog Output setting.