CHAPTER 10
Troubleshooting
119
Known issues
The following issues are known (FW version 1.10.08) and work arounds
are advised. In the second column (instr.) the type of instrument to which
the issue applies is indicated. SCC = Single Cell control (standard prod-
uct), DCC = Dual Cell control (option).
#
Instr.
Description
Work around
1
DCC
Not all pulse mode pa-
rameter (E4, E5, t4 and
t5) of sensor board 1 will
be updated to sensor
board 2 in enslave mode
(CONFIG menu).
Set these specific parameters for
board 2 in the PULSE
2
,
SETUP3
menu
2
SCC,
DCC
Maximum compensation
current values (Max-
Comp) shown in the
SETUP menu are incor-
rect for the 200 nA and
500 nA range.
The maximum current which can
be compensated in these ranges
is 25 µA. See table II and III in
Chapter 3 for an overview of the
maximum compensation for all
selectable ranges in the DC and
pulse mode.
3
SCC,
DCC
In Pulse mode - 2 the
0.005, 0.002 and 0.001
Hz are selectable from
the method window of the
CDS control drivers (Dia-
logue Elite, Chromeleon),
but must not be used. It
can lead to noise and
regular spikes in the
chromatogram.
Select an ADF filter setting in the
range of 0.5
– 0.01 Hz for the
Pulse mode - 2.
4
DCC
Related to issue #3: If
one of the cells is oper-
ated in Pulse mode - 2,
and the other in either DC
or pulse mode: do not
use an ADF filter setting
of 0.005, 0.002 and 0.001
Hz on both cells. These
filter settings can affect
the performance of the
cell operated in Pulse
mode - 2.
Select an ADF filter setting in the
range of 0.5
– 0.01 Hz for both
cell 1 and 2.
Summary of Contents for ROXY Exceed
Page 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Page 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Page 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Page 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Page 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Page 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Page 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Page 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Page 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Page 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Page 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Page 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...