Chapter 7
Optimization of working potential
81
In the example below a half scan is shown at a flow rate of 10 µL/min of a
20 µM Serotonin (5-HT) standard in mobile phase.
Fig. 40. Scan (cycle: half) of a 20 µM Serotonin in mobile phase at a
glassy carbon working electrode. Scan speed 20 mV/s.
The following procedure is advised to record scanning voltammogram of
analytes:
•
Use a voltammetry set-up as shown in figure 48 (preferably) in
combination with the Dialogue Elite software and a syringe pump
for direct infusion of analyte in the flow cell.
•
Prepare a solution of the pure compound dissolved in (preferably)
the HPLC buffer with a concentration in the range of 10-100 µM.
•
Set the lower (E1) and upper potential (E2), the scan rate (Spd),
range and scan cycle
in the ‘SCAN SETUP’ menu or under the
If Dialogue is not used, connect an A/D converter to the analog output
of the potentiostat to record the cell current. Set the A/D converter
sampling rate to 1 Hz. This is the same frequency as the voltage steps
during the scan. If a higher sampling frequency is chosen a typical
stepwise pattern may appear. Note that with such set-
up only ‘I versus
t’ curves can be obtained.
Summary of Contents for ROXY Exceed
Page 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Page 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Page 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Page 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Page 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Page 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Page 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Page 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Page 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Page 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Page 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Page 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...