Fig. 50. Dialogue Elite Events tab with an example of two programmed
events switching the valve to Inject (t = 0.02 min) and back to Load (t =
0.04 min).
ANALOG DATA connector
The ROXY Exceed is equipped with an analog data output connector to
provide the measured signals in millivolts (mV) for users who work stand-
alone without PC control with the instrument. The ANALOG DATA con-
nector (D9-female) can either be connected to an X-Y recorder or A/D con-
verter. The ANALOG DATA output can supply either a non-manipulated
signal directly from the I/E converter, or the processed data signal by the
CPU of the ROXY Exceed. The type of output can be selected from the
CONFIG screen by setting the parameter Vout source to either DAC or I/E
(ROXY Exceed only, for Exceed use Dialogue Exceed to set value).
T e m p
=
3 5 °
C
C o n t
r
a s t
=
2 0
C O N F I
G
V a l
v e =
P r
e s e n t
V o u t
s o u r
c e =
D A C
I
D 1
m a s t
e r
=
n o
P R E V
I
O
Summary of Contents for ROXY Exceed
Page 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Page 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Page 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Page 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Page 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Page 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Page 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Page 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Page 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Page 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Page 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Page 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...