Chapter 5
Operation modes & Parameters
61
Table IV. DC mode filter setting and corresponding data rate.
Filter setting DC mode (Hz)
Data rate (Hz)
RAW
100(default), 50, 20, 10, 5, 2, 1
10
100
5
50
2, 1
20
OFF
10
0.5, 0.2, 0.1
10
0.05
5
0.02
2
0.01, 0.005, 0.002, 0.001
1
Filter OFF is also a special case. The data rate is fixed to 10 Hz, and the
data is not filtered. Setting OFF is therefore the same as RAW at 10 Hz.
Pulse mode
In the pulse mode the working electrode is dynamically and continuously
regenerated by a series of potential steps in a cyclic manner. Data is pro-
cessed differently, and the data rate is defined by the total duration of the 5
potential steps in a pulse: t1 + t2 + t3 + t4 + t5. Usually the typical pulse
duration is between 0.5 and 2s (data rate between 2
– 0.5 Hz). Filter set-
tings in the pulse mode are therefore selectable between 0.5 and 0.001
Hz, and OFF. In Pulse mode 2 filter settings in the range of 0.5 and 0.01
Hz are selectable.
The scan mode has no filter at all.
Summary of Contents for ROXY Exceed
Page 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Page 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Page 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Page 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Page 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Page 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Page 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Page 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Page 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Page 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Page 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Page 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...