Anatel A643a - Calibration and Validation
95 of 220
RP
S -
J
u
ne
20
07
- E
d
it
io
n 1
2
Anatel
Operator Manual
6.5
System Suitability
System suitability provides a confirmation of the performance and efficiency of Anatel’s on-line
Total Organic Carbon instrumentation. The procedure compares the average of a series of
TOC measurements from an easy-to-oxidize working standard (in the form of sucrose) to the
average of a corresponding series of measurements of a hard-tooxidize system suitability
solution (in the form of 1,4-benzoquinone). The response efficiency (RE) should be within 85%
to 115%.
Note:
Refer to Standard Operating Procedure (SOP) #643–2/2A for complete instructions on
performing this procedure.
To determine system suitability:
1)
With the desired Analyzer selected in any View, press the
Manual Key
to access its
menu selections.
2)
Use the
Up
and
Down Keys
to specify Calibrations.
3)
Press
Enter
to display its submenu.
4)
Use the
Up
and
Down Keys
to specify
Suitability
. If the chosen Analyzer is connected
to an A-Net network on which other instruments reside, the analyzer will prompt to
confirm that it is the correct instrument by verifying its serial number. The Analyzer will
begin a temperature stabilization process, its duration depending on the sample water
temperature. The actual time required is displayed and counted down.
CAUTION:
In a networked environment (see
“Network Installation” on page 39
), it is important to confirm
the instrument’s Serial Number before proceeding. Operation of the wrong Analyzer could
result in water leakage from the sample vessel when it is removed during this procedure.
5)
Press
Enter
to continue and wait while the sample temperature stabilizes and the
sample line is drained. Enter the certified sucrose and 1,4-benzoquinone standard
values from the
Certificate of Analysis
in the System Suitability Kit at the prompt.
6)
Press
Enter
to enable the Controller’s Edit Mode and the flashing block cursor
becomes an underscore.
Fig 6-24 :
System Suitability
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......