The following limits apply to each of the Anatel A643a’s Calibration tests. If the limits are
exceeded, a message is logged and printed, and the results of the test will not be saved.
Any questionable problems should be referred to Anatel and the following information should
be included in any correspondence with the factory:
•
The instrument’s serial number.
•
Sample resistivity and temperature.
•
Current and historical TOC data.
•
Current and historical alarm codes.
•
A Setup Printout.
•
Process conductivity or resistivity and temperature.
•
Current and historical calibration, validation and system suitability results.
•
Any other significant changes in operating conditions.
If existing problems warrant equipment return, contact the Anatel Service Department at:
Hach Ultra Analytics, Inc.
5600 Lindbergh Drive
Loveland, Colorado 80538 U.S.A.
Do not
return any equipment to Anatel without a return authorization.
CAUTION:
Drain all water from the Analyzer when taking the instrument out of service (see
). Residual water will freeze inside the unit, expand and damage the
measurement cell. Such damage is time-consuming to repair and is not covered by the
warranty.
Conductivity
Calibration Test
The meter measurement must meet a tolerance of ±0.1 µS/cm. The
cell constant also must not deviate from the previous calibration by
more than 5%, or the factory calibration by more than 10%.
TOC Calibration Test
The percent change in slope must not deviate from the factory
calibration by more than 15%.
System Suitability
Test
If the calculated response efficiency is above 115% or below 85%,
the Analyzer fails the test. A
Code 44: Sys Suitability
is reported
every eight hours until a new system suitability is successfully
performed.
TOC Validation Test
The recognized response must be within ±10% of the given standard
value. A message is logged if the instrument fails the validation test.
Voice: 800-866-7889
+1-541-472-6500
FAX: 1-970-663-9761
Support Hot Line: 800.866.7889
541.472.6500
E-mail: [email protected]
Website: www.hachultra.com
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......