
Anatel A643a Maintenance - Anatel A643a
R
PS
-
June
20
07
-
Ed
it
io
n
12
Operator Manual
Anatel
The following alarm messages also may indicate a deteriorated UV Lamp, requiring that it be
replaced:
•
Code 31: See Guide
•
Code 32: See Guide
•
Code 53: UV Lamp Bad
These messages continue until the UV Lamp is replaced and the Anatel A643a’s internal
counter is reset.
13.4.1 UV Detect Technology
An A643a analyzer equipped with UV Detect will automatically monitor the output of the UV
lamp and alert when the UV output is indicative of a failing lamp or if the lamp has failed
completely and can no longer support proper oxidation. Code 52 indicates a failing lamp.
Code 53 indicates a failed lamp.
Failing Lamp Indications in TOC Analysis Mode:
When Code 52 is indicated by UV Detect, the analyzer UV lamp should be replaced as soon as
possible. UV Detect will generate a Code 52 when the UV output from the lamp has reached a
level that indicate a failing lamp. At this level of UV output the analyzer will continue to oxidize
adequately, but lamp life is extremely limited.
When a failing lamp is detected the analyzer will display
Code 52 UV Lamp
on the C80
controller alarm status screen and the associated LED for the analyzer will blink Red. If a
printer is connected to the analyzer
Code 52 UV Failing
will be printed following each TOC
analysis, and the following message will be printed every eight hours - Urgent: U.V. lamp needs
replacement consult your manual or call the Anatel Service Dept. @ 800-866-7889.
Note:
Under Code 52 the analyzer will continue to perform TOC analysis, but the lamp life is limited
and should be replaced as soon as possible.
Bad Lamp Indications in TOC Analysis Mode:
When Code 53 is indicated by UV Detect the analyzer lamp has failed and must be replaced.
Code 53 is an indication that the UV lamp has failed and no longer supports proper analyzer
operation.
Under a Code 53 condition the lamp is no longer outputting adequate levels of UV light to
support proper sample oxidation. TOC analysis is no longer supported and the C80 display will
display
53-UV Lamp Bad
.
When a bad lamp is detected the analyzer will display
Code 53 UV Lamp
on the C80 controller
alarm status screen and the associated LED for the analyzer will blink Red. If a printer is
connected to the analyzer
Code 53 UV Lamp Bad
will be printed following each attempt at a
TOC analysis, along with the following message - Urgent: U.V. lamp needs immediate
replacement, consult your manual or call the Anatel Service Dept. @ 800-866-7889.
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......