General Information
Anatel A643a Analyzers require periodic maintenance in order to ensure the continued
accuracy and reliable performance of the instrument. The UV lamp has a limited life and should
be replaced every six months. It is recommended that the measurement cell have an annual
calibration verification. The lithium battery has a service life of 3–5 years.
Maintenance procedures may be performed either by the user or by Anatel customer service
personnel. The most common maintenance procedures are UV lamp replacement, cleaning
the air filters, lithium battery replacement and printer maintenance. In most instances, the user
can perform these procedures in the field. If problems persist after performing maintenance or
repair procedures, contact the Hach Ultra Customer Service department at 541.472.6500.
Should it be necessary to return Anatel equipment to the factory for maintenance or repair,
obtain a Return Materials Authorization (RMA) number. At that time furnish the Model and
serial number, and any additional relevant information. Ship the instrument(s) to
Hach Ultra,
Attn: RMA #xxxx.
CAUTION:
Purge all water from the Analyzer when taking the instrument out of service. Residual water
can freeze inside the unit, expand and damage the measurement cell. Such damage is
timeconsuming to repair and is not covered by the warranty.
Anatel also offers service and extended warranty options to cover scheduled maintenance,
calibration and repairs. In cases where returning the Analyzer to the factory is impractical, on-
site maintenance and/or verification can be performed by an Anatel technician, or a certified
instrument may be rented. Repairs are performed only at Hach Ultra Service Centers due to the
specialized equipment and parts required.
13.2
Self-Cleaning Mode
It is possible to contaminate an Anatel A643a Analyzer’s measurement cell on any water
system or after long-term storage or inactivity. This condition is indicated by suddenly changing
or erratic TOC or resistivity readings. Extended self-cleaning using the ultraviolet lamp to
oxidize contaminants in the cell is the recommended procedure to alleviate such problems. In
the Self-Clean Mode, the Analyzer’s sample valve is opened to allow water to flow through the
cell. The UV lamp is turned on to oxidize any organic contaminants, which subsequently are
flushed from the instrument by the sample flow.
Note:
The Anatel A643a Analyzer must have the sample vessel attached and be on-line with sample
pressure so that there is flow through the instrument.
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......