Calibration and Validation - Anatel A643a
R
PS
-
June
20
07
-
Ed
it
io
n
12
Operator Manual
Anatel
6.2
TOC Calibration
Note:
Refer to Standard Operating Procedure (SOP) #643–1/1A for complete instructions on
performing this procedure.
Calibration ensures the accuracy of the TOC values obtained by the Anatel A643a Analyzer.
TOC values are determined by measuring the change in conductivity of a high-purity water
sample as its organic compounds are oxidized to carbon dioxide inside the instrument’s
analysis cell. The confirmation procedure subtracts the average of a series of background TOC
measurements from the average of a series of working standard measurements.
Linear regression is performed on the data and a correlation coefficient (R) for the analysis is
reported. The instrument automatically determines whether the calibration passes or fails
based on the factory calibration. If the absolute value of the percent slope change (%
SC
) of the
new calibration is
≤
15% of the factory calibration, the instrument indicates that the new
calibration has passed. If the result is >15% of the factory calibration, the Analyzer indicates
that the new calibration has failed.
Note:
It is recommended the correlation coefficient, a user-accepted (R) value, be
≥
0.990. The
calibration will not fail if the R value is
≤
0.990, only if the % slope change is not within the
specified limits.
To perform a TOC calibration:
1)
With the desired Analyzer selected in any View, press the
Manual Key
to access its
menu selections.
2)
Use the
Up
and
Down Keys
to specify
Calibrations
.
3)
Press
Enter
to display its submenu.
4)
Use the
Up
and
Down Keys
to specify
Calibrate
.
5)
Select
TOC Calibrate
and press
Enter
. If the Analyzer is connected to an A-Net
network on which other instruments reside, confirm that it is the correct instrument by
verifying its serial number. The Analyzer will begin a temperature stabilization process,
its duration depending on the sample water temperature. The actual time required is
displayed and counted down.
Fig 6-1 :
TOC Calibration
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......