Anatel A643a - Calibration and Validation
81 of 220
RP
S -
J
u
ne
20
07
- E
d
it
io
n 1
2
Anatel
Operator Manual
6
Calibration and Validation
6.1
General Information
The Anatel A643a is equipped with on-demand calibration and validation capability to ensure
the reliability of its results. Calibration assures the accuracy of instrument’s readings at regular
intervals; validation ensures the accuracy of the calibration by “challenging” it with the analysis
of a known standard. Calibration and validation data are available for viewing via a C80
Controller or through a local or network printer.
Note:
The
Anatel A643a
is calibrated at the factory before shipment. It is not necessary to perform a
TOC Calibration prior to placing the instrument into service.
Proper calibration of the Anatel A643a is critical to maintaining optimum instrument
performance. There are several approaches to calibration to run unknown samples and
accurately report the results. Calibration and validation involve running analyses of one or more
standards and blanks in order to determine instrumental response:
Each procedure is prompted by a series of menus and all analyses are performed
automatically. The necessary materials are available through Anatel in prepackaged
Reagent Kits.
Note:
These are intended as general instructions on the performance of the calibration and
validation procedures. Refer to the specific Standard Operating Procedure (SOP) for detailed
instructions.
TOC Calibration
Select from up to three levels (250, 500 and 750 ppb) of a
sucrose standard. The procedure also allows up to three
analysis replicates per standard for maximum accuracy.
TOC Validation
Perform either a single-point or multipoint validation of the
TOC calibration using using up to three levels of sucrose
validation standards.
Conductivity Calibration
Performed according to the guidelines established in
USP Method <645>. The procedure uses a NIST-
traceable calibrated resistor for the determination of
meter accuracy and a NIST-traceable 100 µS/cm
conductivity standard for cell constant verification.
System Suitability
Performed according to the guidelines established in
USP Method <643> using prepackaged 500 ppb sucrose
and 500 ppb 1,4-benzoquinone standards.
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......