After specifying a given ppb concentration level, press
Enter
to highlight the
concentration value. Using the
Up
and
Down Keys
, enter the actual TOC value for the
given standard concentration as it appears on the
Certificate of Analysis
and press
Enter
to select the final value. Repeat this procedure for all calibration standard levels
selected.
9)
Also highlight the
# of reps
and specify the number of replicates to be performed at
each level. Press
Up
and
Down Keys
to toggle through the available selections (1–3).
Typically, the more replicates, the more accurate the calibration results. With the
desired number replicates displayed, press
Enter
.
10)
Press
Enter
. The analyzer will asked if the “C of A” values have been entered. If not,
press
Esc
to enter the “C of A” values: once they are properly entered, press
Enter
to
continue.
11)
If any changes must be made to the standard concentrations or the number of
replicates, press
Esc
followed by
Enter
to abort the procedure. Immediately press
Enter
at the
TOC Calibrate Menu
to re-enter the
Standard Selection Menu
.
12)
Remove the sample vessel and install the blank vial at the prompt. Press
Enter
and its
contents are sampled and analyzed. The results of each of the analysis replicates are
reported.
Fig 6-3 :
Calibration Purge Mode
Fig 6-4 :
Calibration Mode
Calibration
Vial
Needle
Valve
Check
Valve
15 Micron
Filter
Bypass
Check
Valve
Sample
Valve 2
Calibration
Valve
Pump
Sample
Valve 1
Analysis
Cell
Thermistor
Sample
Needle
Check
Valve
Air
Filter
Air Inlet
Drain
Process
Water
Inlet
Calibration
Purge Mode
Note:
Arrows indicate flow path.
Calibration
Vial
Check
Valve
Air
Filter
Needle
Valve
Check
Valve
15 Micron
Filter
Bypass
Check
Valve
Sample
Valve 2
Calibration
Valve
Pump
Sample
Valve 1
Analysis
Cell
Thermistor
UV Lamp
Sample
Needle
Drain
Air Inlet
Process
Water
Inlet
Calibration
Mode
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......