Anatel A643a - Analog Outputs
129 of 220
RP
S -
J
u
ne
20
07
- E
d
it
io
n 1
2
Anatel
Operator Manual
5)
Press
Esc
repeatedly to exit this function and return to the normal display.
9.3
Optional Conductivity/Resistivity & Temperature Outputs
Note:
Resistivity DAC: P/N FG2006702
Temperature DAC: P/N FG2006802
The necessity for an analog output to report the Anatel A643a’s conductivity (or resistivity) and
temperature values is addressed through external Digital-to-Analog Converter (DAC) Modules.
A separate DAC Module is required for each desired analog output. Anatel supplies pre-
programmed, prewired DAC Modules to forward this processed information to a data
acquisition system, remote meter or other analog input device. Software setup of the External
DAC Module includes enabling the Analyzer output, defining its range as well as its zeroscale
and full-scale values, and the signal that is to be output in the event of an alarm.
9.3.1
Hardware Setup
Note:
Refer to the manufacturer’s manual for complete instructions on installing and using external
DAC Modules.
External DAC Modules are connected to the Anatel A643a at its DIAGNOSTICS Port. This
RS232C serial interface communicates at 1200 baud, 8 data bits, 1 stop bit and no parity.
Three- or five-conductor cable is adequate for this analog interface (AWG 24 is recommended).
Two DAC Modules can be daisychained to obtain both conductivity (or resistivity) and
temperature data and can be positioned up to 50 feet from the associated Analyzer. Anatel
provides five feet of cabling.
CAUTION:
Always disconnect the instrument from its power source before attempting to access internal
components.
Fig 9-17 :
DAC Calibration Output Screen
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......