Analog Outputs - Anatel A643a
R
PS
-
June
20
07
-
Ed
it
io
n
12
Operator Manual
Anatel
9.2.2.4
DAC Error Output
Note:
Default: Minimum Output
The user may select from among three states to be output when a condition is encountered
which causes the Anatel A643a to abort its current analysis. This fixed signal is global for all
analog outputs and can be used externally to report an alarm condition.
Note:
In the Calibrate mode, the analog output reflects on-line values. A zero-scale analog output
indicates the instrument is off-line; a full-scale output indicates an error condition.
To specify the
DAC Error Output
:
1)
With the Analyzer’s
Analog Setup Menu
displayed, use the
Up
and
Down Keys
to
specify
DAC Modes
.
2)
Select
DAC Err Output
and press
Enter
to display its parameter screen.
3)
Use the
Up
and
Down Keys
to specify the desired setting:
4)
Press
Esc
to return to the
DAC Modes Menu
. Select
DAC Cal Output
to enable that
feature, if desired. When “Active,” an analog signal proportional to each analysis result
is sent to an external DAC Module during calibration.
1 SENSOR NAME
X
DAC Err Output:
DAC Cal Output:
Fig 9-15 :
DAC Error Output Menu
Minimum Output
Transmits the Analyzer’s corresponding Zero-Scale TOC
value.
Last Output
Transmits the Analyzer’s last reading.
Maximum Output
Transmits the Analyzer’s corresponding Full-Scale TOC value.
Fig 9-16 :
DAC Error Output Screen
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......