CAUTION:
Some instrument surfaces may be hot if the sample water is hot. Allow surfaces to cool before
touching.
CAUTION:
In a networked environment (see
“Network Installation” on page 39
), it is important to confirm
the instrument’s Serial Number before proceeding. Operation of the wrong Analyzer could
result in water leakage from the sample vessel when it is removed during this procedure.
Note:
The Anatel A643a TOC Analyzer was designed to be calibrated by the user. Anatel provides
pre-packaged reagents to assist the user in simplifying the calibration process and to easily
comply with USP TOC Method <643>. If desired, reagents can be prepared by the user for use
with the A643a TOC Analyzer. User-prepared reagents will not affect or alter the calibration of
the instrument. User prepared standards can be introduced into the analyzer using a
container/closure system that includes a 50 mL glass vial (VWR cat #16171341), a 20 mm
teflon-coated stopper (VWR cat #16171561) and a 20 mm crimp-top aluminum seal (VWR cat
#16171829). The user is responsible for the traceability and USP compliance of all reagents
that are not pre-packaged.
6)
Press
Enter
to continue and wait while the sample line is drained. The remaining time
is counted down on the display.
7)
Specify the calibration standard levels to be performed. All three available ppb
concentration levels are enabled by default, indicated by the checkmark (
3
) preceding
them. Disable any unwanted levels by pressing the
Up
and
Down Keys
to select the
desired concentration, then pressing
Enter
to edit and
Up
and
Down
toggle it on and
off. At least two levels must be selected, however.
Note:
Calibration requires at least two Replicates at two levels of Standards. The concentration
of Standards used should bracket the concentration of the samples to be analyzed. Three
Standards with three Replicates each is recommended for the best accuracy of calibration.
Fig 6-2 :
Calibration Standard Criteria
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......