ADM & Sub Aspida
User Manual
Int. Approved
Document ref: PSA-811-22
March 2020
Page 8 of 45
Copyright © 2017 Analox Ltd. All Rights Reserved.
Commercial in Confidence
3
Aspida devices overview
The Analox Aspida
is a purpose designed compact portable gas monitor capable of continuous
monitoring of both carbon dioxide (CO
2
) and oxygen (O
2
).
Each instrument in the Aspida family gives clear audible and visible warning of potentially
dangerous gas levels. A high-resolution Organic Light-Emitting Diode (OLED) display shows clear,
live gas levels in all light conditions.
The Analox Aspida is housed in a robust, IP65 splash proof enclosure. The instrument operates
using rechargeable battery technology, allowing it to run for more than 12 hours continuously
between charges. An Analox Aspida allows for easy replacement of the rechargeable batteries with
standard, AA type non-rechargeable batteries in circumstances where recharging is not possible.
Sub Aspida overview
The inclusion of a pressure sensor in the Sub Aspida provides accurate, pressure compensated O
2
and CO
2
readings across 800 to 1200mbar pressure.
The inclusion of a belt clip attachment allows for secure and comfortable attachment of the Sub
Aspida to clothing for portable, personal protection.
Table 1
Sub Aspida factory default settings
Parameter
Setting
O
2
Alarms
%Vol Version
mbar ppO
2
versions
18%
19.5%
23%
180 mbar ppO
2
195 mbar ppO
2
230 mbar ppO
2
CO
2
Alarms
%Vol Version
mbar ppCO
2
versions
0.5%
1.5%
4%
0.5% TWA Alarm
5 mbar ppCO
2
15 mbar ppCO
2
40 mbar ppCO
2
5 mbar ppCO
2
TWA Alarm
Alarm Latch State
Non-latching
Alarm Muting State
Mutable
Man-down State
Disabled
Logging Rate
30 seconds
ADM Aspida overview
The ADM Aspida offers continuous monitoring of O
2
, CO
2
or dual gas. The ADM Aspida can be
retrofitted with a faceplate allowing for panel mounting, see section 7.1 for the faceplate kit part
number.
Table 2
ADM Aspida factory default settings
Parameter
Setting
O
2
Alarms
%Vol Version
18%
19.5%
23%
CO
2
Alarms
%Vol Version
0.5%
1.5%
4%
Alarm Latch State
Non-latching
Alarm Muting State
Mutable
Man-down State
Disabled
Logging Rate
30 seconds