ADM & Sub Aspida
User Manual
Int. Approved
Document ref: PSA-811-22
March 2020
Page 31 of 45
Copyright © 2017 Analox Ltd. All Rights Reserved.
Commercial in Confidence
Figure 22
Calibration progress
15]
When the sampling process has complete, an audible alert will be given, and a
calibration confirmation screen will be shown as below.
Figure 23
Calibration failed
Figure 24
Calibration passed
16]
If the calibration passed, then the tick icon will be shown. The new calibration
information will then be stored to the device memory and the device is ready for
use. Press the
button to confirm the operation and return to the main screen.
17]
If a stable carbon dioxide reading cannot be detected, then the confirmation
screen will show a cross icon. Press the
button to acknowledge the failed
calibration. A failed calibration will leave the previous gas calibration unchanged.
To attempt the calibration again repeat steps 3
–
8.
Once calibrated successfully, the calibration due date for the sensor will be reset.