ADM & Sub Aspida
User Manual
Int. Approved
Document ref: PSA-811-22
March 2020
Page 27 of 45
Copyright © 2017 Analox Ltd. All Rights Reserved.
Commercial in Confidence
6.23
Troubleshooting
Symptom
Possible cause
Action
Device does not switch on
No batteries, or incorrectly
orientated batteries
Check that batteries
are inserted
correctly
Device does not switch on
Batteries are flat
Check that the
batteries are
inserted correctly
and are
rechargeable NiMH
then perform a
device charge.
Batteries will not charge (
shown flashing)
Batteries may not be
rechargeable type
Check that the
batteries fitted are
the rechargeable
NiMH batteries
supplied with the
Aspida
Batteries will not charge (
shown flashing)
Aged batteries causing a
charge fault
Contact supplier for
information about
replacing
rechargeable
batteries
Batteries will not charge (
shown)
No batteries fitted
Check that
rechargeable
batteries are fitted
Battery charge completes but
battery life is short
Battery ageing reduced battery
capacity
Contact supplier for
information about
replacing
rechargeable
batteries
Flashing symbol underneath
gas reading
The sensor is due a
recalibration
Refer to section 7.2
of this manual for
instructions
Flashing symbol underneath
oxygen gas reading
The oxygen sensor is due to be
replaced
Refer to section 7.5
of this manual for
instructions