ADM & Sub Aspida
User Manual
Int. Approved
Document ref: PSA-811-22
March 2020
Page 21 of 45
Copyright © 2017 Analox Ltd. All Rights Reserved.
Commercial in Confidence
6.10
Global alarm options
There are three configurable options which affect all of the Analox Aspida
’
s alarms. Muting,
latching and quiet options are set using the Aspida configuration software. See the Aspida
configuration software manual for instructions.
6.11
Alarm set-points
The alarm set-points and direction can be changed using the Aspida configuration software. See
the Aspida configuration software manual for instructions. An ADM Aspida also has the facility to
change the alarm set-points on the device as per section 6.11.1
6.11.1
On device alarm set-points (ADM Aspida only)
The below procedure is the same for both the CO
2
and O
2
alarm setpoints.
1]
Press the
button to access the device menu.
2]
Use the
button multiple times to highlight the
icon or the
icon
(initially off the display to the right) and press
.
Figure 6
Alm CO
2
or Alm O
2
selection
3]
Using the
and
. Buttons select the relevant
,
or
icon and then
use the and icons to adjust the displayed value to the required alarm value.
Figure 7
Selecting the alarm setpoint
4]
When correctly set, confirm the alarm value using the
icon and
. Button.
Figure 8
Example CO
2
Alm 1 set-point
Figure 9
Example O
2
Alm 1 set-point
Summary of Contents for ADM Aspida
Page 2: ......