background image

 

 

Welcome to visit ATO.com 

Tel: 800-585-1519      Email: [email protected]    Web

www.ato.com

 

User Manual- YT-1200H-S4 

 

Thanks for purchasing our product. Please read the user manual carefully before using. 

 

Keys Operation: 

 

ON/OFF:

 long press for 3s to turn on/off the device. In detection mode, short press to check battery 

power, and press again to go back to detection mode.  In parameter setting mode, short press to 

confirm or save settings. 

 

UP:

 in detection mode, long press the UP button to enter into menu, and in parameter setting mode, 

short press this button to move to upper item. 

 

DOWN:

 in detection mode, long press to shift the gas concentration unit quickly, but it is restricted to 

PPM and mg/m3. In parameter setting mode, short press it to go back to move to next item. 

 

(Note: The long press function only work in detection interface. When you change or reset any data, please 

remember to click “ON/OFF” button to confirm and save the settings) 

 

 

Power On 

Press the “ON/OFF”  for three seconds, the detector will be turned on.  The screen displays the following 

interface successively: Brand and Logo, main parameter interface, initialing countdown. The countdown would 

be 30~60 seconds, which is to make sure that the sensors are fully activated. After 30 seconds, the device 

enter normal detection mode.   

 

 

 

Zero Calibration 

Zero calibration must be done in clean air. Zero calibration is needed when there is zero shift happens. 

 

Put the detector in clean air for 3-5 minutes, under the detection mode, press the “UP” and “DOWN” buttons at 

the same time for three seconds, the data of all channels will reset to zero automatically.   

 

Note: Zero calibration of oxygen detector, nitrogen detector and carbon dioxide detector (target gases are 

component of the air), the data will be defaults to 20.9%VOL (O2), 79.1%VOL (N2), 400ppm (CO2). Do not do 

the zero calibration in not clean air or when there is no zero shift happens. 

 

 

www.ato.com

 

Summary of Contents for YT-1200H-S4

Page 1: ...et any data please remember to click ON OFF button to confirm and save the settings Power On Press the ON OFF for three seconds the detector will be turned on The screen displays the following interface successively Brand and Logo main parameter interface initialing countdown The countdown would be 30 60 seconds which is to make sure that the sensors are fully activated After 30 seconds the device...

Page 2: ... until the reading back to zero before powering it off and do the zero point calibration next time before using it If the reading maintained at full scale user should send the device back to manufacturer or agent for repair or replacing the sensor Alarm light Silicon rubber case LCD USB charging port UP DOWN ON OFF Sensor air chamber Note If detector is used for the gas cylinder detection consider...

Page 3: ...there will be sound light low battery alarm There will be sound and light every 5 seconds to remind the user to charge the device in time When the battery is used up the device will be turned off automatically Alarm Setting In detection model long press UP button to enter the menu choose the Alarm point and start to set the alarm value There are four options under the alarm setting mode The first ...

Page 4: ... Prepare the standard gas before starting the target point calibration Choose standard gas press ON OFF button to enter There will show please input standard gas at the top of the LCD and at the bottom there are gas type calibrate value VA current concentration AD and ESC or SET Change the target concentration to the standard gas concentration by press UP and DOWN after the value is stable press t...

Reviews: