Process of Over-range incorrect operation 

Users should avoid using gases which exceed the measure range to impact the sensor, because it will affect 

the lifespan and sensitivity of the sensor, even, "poison" the sensor. If there is any over-range incorrect 

operation makes the detector displaying an concentration at a large reading, the remedy is to take the device 

out of the environment immediately and put it in clean air for over half an hour, and then observe the reading, if 

it keeps going down, then wait until the reading back to zero before powering it off, and do the zero point 

calibration next time before using it. If the reading maintained at full scale, user should send the device back to 

manufacturer or agent for repair or replacing the sensor. 

 

 

 

 

 

 

 

                                                                 

                                                  

                                                  

  — Alarm light 

 

               

 —Silicon rubber case 

 — LCD 

 — USB charging port 

 — UP 

 — DOWN 

 —ON/OFF 

— Sensor air chamber 

 
 
------------------------------------------------------------------------------------------------------------------------------- 

Note 

*If detector is used for the gas cylinder detection, considering the high pressure in the cylinder, please use a 

pressure regulator, and keep it as the below data: 

Flow rate: 800mL~1L/minute 

Pressure: 0.1MPa or 1Bar 

* Working temperature: -20~50 degree C 

* Working humidity: 0-95%RH, no condensing 

* Please put the detector in clean air for about 5 minutes and waiting for the data falling down to normal before 

turning it off after measurement.

 

www.ato.com

 

Summary of Contents for YT-1200H-S4

Page 1: ...et any data please remember to click ON OFF button to confirm and save the settings Power On Press the ON OFF for three seconds the detector will be turned on The screen displays the following interface successively Brand and Logo main parameter interface initialing countdown The countdown would be 30 60 seconds which is to make sure that the sensors are fully activated After 30 seconds the device...

Page 2: ... until the reading back to zero before powering it off and do the zero point calibration next time before using it If the reading maintained at full scale user should send the device back to manufacturer or agent for repair or replacing the sensor Alarm light Silicon rubber case LCD USB charging port UP DOWN ON OFF Sensor air chamber Note If detector is used for the gas cylinder detection consider...

Page 3: ...there will be sound light low battery alarm There will be sound and light every 5 seconds to remind the user to charge the device in time When the battery is used up the device will be turned off automatically Alarm Setting In detection model long press UP button to enter the menu choose the Alarm point and start to set the alarm value There are four options under the alarm setting mode The first ...

Page 4: ... Prepare the standard gas before starting the target point calibration Choose standard gas press ON OFF button to enter There will show please input standard gas at the top of the LCD and at the bottom there are gas type calibrate value VA current concentration AD and ESC or SET Change the target concentration to the standard gas concentration by press UP and DOWN after the value is stable press t...

Reviews: