Calibration 

Note: Calibration needs to be done by professionals and with necessary accessories and standard gas. 

Password is needed. (If you need password, please contact the manufacturer to get it) 

 

In detection mode, long press the “UP” button, choose the CALIBRATION. There are five options in the 

calibration interface: 

1.  Channel option: you need to choose the right channel/gas type before calibration, there will be 4 channels 

for the 4 gases multi gas detector. 

2.  Zero calibration option: put the detector in clean air for 3-5 min, then press ”ON/OFF” button, wait for the 

value become stable, press ”ON/OFF” again, the zero calibration is finished. You can also do the zero 

calibration in the detection mode, press the “UP” and “DOWN” buttons for three seconds, the data of all 

channels will reset to zero automatically. 

 

3.  Target point calibration:   

Prepare the standard gas before starting the target point calibration. Choose standard gas, press ON/OFF 

button to enter.   

There will show “please input standard gas” at the top of the LCD, and at the bottom there are “gas type”, 

calibrate value, “VA” (current concentration), “AD”, and “ESC or SET”. Change the target concentration to the 

standard gas concentration by press “UP” and “DOWN”, after the value is stable, press the “ON/OFF”, to finish 

the calibration. 

 

The detailed process is as below: 

1.  Connect the detector with standard gas cylinder by tube. 

2.  Enter the standard gas calibration interface. 

3.  Change the target point value to the standard gas value. 

4.  Release the standard gas to the detector in a flow rate of 400mL/min. 

The value of “current concentration” will become larger gradually, and 

become stable after about 30 seconds. 

5.  When the value of VA and AD is stable, the value of “current 

concentration” will be same with or almost same with the “target point 

value”. Press the “UP/DOWN” button to change the “SET” to 

“CONFIRM”, and press the “ON/OFF” button, the calibration is 

finished. 

6.  Cut off the standard gas supply.   

 

 
 

www.ato.com

 

Summary of Contents for YT-1200H-S4

Page 1: ...et any data please remember to click ON OFF button to confirm and save the settings Power On Press the ON OFF for three seconds the detector will be turned on The screen displays the following interface successively Brand and Logo main parameter interface initialing countdown The countdown would be 30 60 seconds which is to make sure that the sensors are fully activated After 30 seconds the device...

Page 2: ... until the reading back to zero before powering it off and do the zero point calibration next time before using it If the reading maintained at full scale user should send the device back to manufacturer or agent for repair or replacing the sensor Alarm light Silicon rubber case LCD USB charging port UP DOWN ON OFF Sensor air chamber Note If detector is used for the gas cylinder detection consider...

Page 3: ...there will be sound light low battery alarm There will be sound and light every 5 seconds to remind the user to charge the device in time When the battery is used up the device will be turned off automatically Alarm Setting In detection model long press UP button to enter the menu choose the Alarm point and start to set the alarm value There are four options under the alarm setting mode The first ...

Page 4: ... Prepare the standard gas before starting the target point calibration Choose standard gas press ON OFF button to enter There will show please input standard gas at the top of the LCD and at the bottom there are gas type calibrate value VA current concentration AD and ESC or SET Change the target concentration to the standard gas concentration by press UP and DOWN after the value is stable press t...

Reviews: