ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 33 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
7]
Connect the sensor wires of the new cell PCB to the corresponding red, blue and yellow
terminals of connector J3.
8]
Feed the PCB back through the bulkhead and refit the lid.
Using a terminal screwdriver press firmly
down on the terminal clamp button to
open the clamp.
While still pressing down on the terminal
clamp button insert the bootlaced wire into
the terminal clamp as shown.
Release the terminal clamp button. The
clamp should now have a firm grip on
the wire. Gently pull on the wire to
confirm that the wire is firmly held. If not
repeat the steps shown until the wire is
held firmly.
Repeat this for the other wire.