ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 27 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
4.2.2
Particulate filter replacement
The particulate filter is used to remove larger particles of VOC’s as
they can after time block the inlet
to the PID sensor, thus making the sensor read erratically or not read at all.
The procedure for changing the particulate filter is as follows:
1]
Unscrew the filter from both of the male and female luers attached to the 6mm
pipework from the inlet manifold.
2]
Reverse the above process to fit a new filter.
4.3
Service requirements
Certain elements of the ACG+ system are recommended to be replaced every six years to ensure optimum
operation.
below shows the maintenance that is required and the scheduled interval for performing the
maintenance.
Table 8 Service schedule of ACG+ components
System component
Maintenance required
Maintenance
interval
VOC sensor
Replace PID cell
5 years
Display module
(See note)
Replace display
6 years
Carbon dioxide sensor
(See note)
Replace sensor
6 years
NOTE:
ACG+ TO BE RETURNED TO ANALOX FOR REPLACEMENT OR
SERVICE
Before attempting to replace any sensors ensure that the ACG+ is switched off and the gas supply has been
isolated. Using a Pozi-drive screwdriver loosen the 4 enclosure lid screws and open the lid.