ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 30 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
5]
Connect the new PID to the PCB connector. Note that it will only connect in one
orientation.
6]
Fit the PID into the housing and rotate anti-clockwise to lock in place.
7]
Refit the PID holder into the sensor bulkhead.
8]
Refit the flow adaptor into the sensor inlet.
NOTE:
FOLLOWING A VOC PID LAMP REPLACEMENT OR
CLEANING, OR, REPLACING THE CELL A FULL CALIBRATION
MUST BE PERFORMED. SEE P0149-801 ACG+ USER
MANUAL FOR FULL DETAILS.
4.3.3
Replacing the PID cell PCB (MEC-233C)
1]
Remove the flow adaptor from the sensor inlet.