ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 16 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
4.1.2
Carbon monoxide cell replacement
The part number for your sensor’s replacement cell is:
9100-2030S
When a replacement cell is received check that the cell has not leaked. The cells are themselves sealed
and do not under normal circumstances present a health hazard however if leakage of the electrolyte
has occurred use rubber gloves and wear chemical splash goggles to handle and clean up. Rinse
contaminated surfaces with water. If contact is made with the electrolyte, please refer to Section
Error! Reference source not found.
1]
Remove the flow adaptor from the sensor inlet.
2]
Use a flat blade screwdriver to pop out the cell holder from the sensor bulkhead fitting.