ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 18 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
NOTE:
FOLLOWING A CARBON MONOXIDE CELL REPLACEMENT
A FULL CALIBRATION MUST BE PERFORMED. SEE P0149-
801 FOR DETAILS. ALSO, A REPLACEMENT NOTIFICATION
RESET SHOULD BE PERFORMED. SEE P0149-801 ACG+
USER MANUAL FOR DETAILS.
4.1.3
Replacing the CO cell PCB (MEC-233B)
1]
Remove the flow adaptor from the sensor inlet.
2]
Use a terminal screwdriver to pop out the sensor holder from the sensor bulkhead
fitting.