PN 9000-133-VE, Rev U
General Wiring and Conduit Requirements
This section describes general wiring and conduit requirements.
• Analyzer wiring conductors must be rated at a minimum of 80 °C. All other
wiring conductor ratings should be for the minimum temperature required
for the equipment being connected to the analyzer, but not less than 60 °C.
• Use only the applicable NEMA-approved conduit fittings or cable fittings
to maintain the NEMA rating for the analyzer enclosure or WDG-V. If not us-
ing a conduit entry, leave the factory NEMA-approved plugs intact. Never
leave any holes unplugged.
• Follow all applicable electrical codes for your location.
• Follow proper grounding, shielding and noise protection practices as
described in this section.
• For all analyzer and signal wiring use twisted-pair cable, 18 to 24 AWG
(American Wire Gauge) (0.82 mm
2
to 0.33 mm
2
), with an overall braided
shield, or twisted-pair cable in rigid metal conduit.
• For AC mains supply wiring, use between 12 and 14 AWG or equivalent
metric between 3.3 mm
2
and 2.1 mm
2
.
• Use the conduit entry point closest to the connections you are making.
Do not add any additional conduit entry holes!
(4) 3/4"NPT
Conduit Entries
Supplied with
Nema Conduit Plugs
42454JE
(2) 1/2"NPT
Conduit Entries
Supplied with
Nema Conduit Plugs
42464JE
SS External
Ground Stud
M5 Ring Terminal
12-10AWG
Figure 2-13.
Conduit entry and ground
stud locations.
2-26 | Thermox
®
WDG-V / VC / VCM / VM
Summary of Contents for WDG-V
Page 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Page 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Page 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Page 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Page 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Page 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...